Particle and Aerosol Research
Vol. 14, No. 4, December 2018, Pages 191-203
|
ISSN : 1738-8716 (Print)
ISSN : 2287-8130 (Online)
|
Development of Drying Systems for Accurate Measurement of Particulate Matter by means of Optical Particle Measuring Instruments
Doo Soo Kang1), Jung Eun Oh1), Sang Yul Lee1), Hee Joon Shin1), Ha Kyung Bong1), Joohyun Choi1), Dae Seong Kim2) *
1)R&D Center, Sentry Inc.
2)Aeris
*Corresponding author.
Tel.: +82-10-6788-8929, E-mail: dskim94@hanmail.net
Received 26 December 2018, Revised 28 December 2018, Accepted 28 December 2018, Available online 31 December 2018
http://dx.doi.org/10.11629/jpaar.2018.14.4.191
Abstracts
IIn this study, we have developed drying systems for reducing the error by humidity on measuring particulate matter (PM) in the ambient air with optical particle measuring instruments. Two types of drying systems were designed: drying systems using heating and dilution methods. In addition, 3 types of drying systems using a heating method were designed: Type A (1 hole), B (3 holes) and C (7 holes). After making them, the laboratory and field tests were carried out to evaluate the developed drying systems. As a result, it was shown that the PM concentrations obtained by PM monitoring devices with drying systems agree well with that of the reference devices. Therefore, it could be concluded that the drying systems can be applied to PM monitoring devices for real-time monitoring of the ambient aerosols.
Keywords
drying system, heating method, dilution method, particulate matter, aerosol measurement
File
10.³í¹®14±Ç4È£(°µÎ¼ö) PAAR-E-201800029 - ÆíÁý2Â÷.pdf
Download